Crude Fiber Tester Operation Procedure

The CXC series crude fiber analyzer uses the principle of acid-alkali digestion (decontamination of the sample under specific conditions, removal of soluble matter by ethanol, depletion of minerals after high temperature ignition). The operating procedure (operation must be carried with thick rubber gloves) is as follows:
1. Take off the three flasks on the top of the instrument first, then close the three valves connecting the rubber tube (the switch and the valve are kept in the vertical state to be closed), then add the prepared acid, alkali and distilled water respectively (should be not less than 2000ml ), Put the cap on the bottle and place it on the heating plate on the top of the instrument.
2. Put the ring on the sealing ring first, then place 1-2g (accurate to 0.0002g) sample in the bowl. Pull the lifting handle up again. At this time, the filter cup holder will drop automatically. Then put the 坩埚 into the 6 filter holders. Press the lifting handles again, then the suction cups will be clamped and the cooking tube will clamp each other (note that when you press the lifting handle, make sure that the bottom cup is sucked and the cooking tube is aligned with each other so as not to damage the glass). Then hand-tighten the 3 pieces of glass to each other to ensure that liquid does not leak out during operation.
3. Connect the water inlet on the left side of the instrument to tap water, then turn on the tap water switch, pay attention to the moderate amount of water. Turn on the power switch first, and then turn on the acid, alkali, distilled water preheat switch. Adjust the preheat voltage knob and turn it clockwise to the maximum. At this time, the voltmeter shows that the voltage is about 220V.
4, when the acid, alkali, distilled water boil, the preheating voltage is adjusted to acid, alkali, distilled water micro-boiling.
5. After the above liquid is slightly boiled, adjust all eight red ball switches below to be horizontally closed. Then insert the valve outlet on the acid bottle rubber tube into the boiling tube, slowly open the valve, and add the acid to The volume in the decontamination tube (from left to right, number 1, 2, 3, 4, 5, and 6 decoction tubes) is 200 ml (about the horizontal position of the middle of the decontamination tube), close the valve, and then Add 2 drops to the digester tube. Turn off the acid preheat switch. Place the heat shield close to the front of the heating box and turn the pressure adjusting knob of the boiling tube to maximum. At this time, the corresponding voltmeter shows about 220V. After the boiled tube, the acid is slightly boiled and then the voltage is adjusted to 150-170V. Left and right, keep the acid slightly boiling, open the timer switch upwards, set the time to 30 minutes (usually adjusted at the factory). After 30 minutes of cooking, the buzzer sounds.
6. Turn the cooking heat switch off, remove the heat shield, turn the pressure adjustment knob counterclockwise, turn on the corresponding valve of the 1-6 glass tube (open the suction filter individually), and then open the suction valve. Finally, open the suction pump control switch to remove the acid. After pumping off, close the suction pump control switch, close the corresponding valve of the glass tube, and finally close the suction filter valve. (If helium is found to be blocked during suction filtration, the suction pump switch can be turned off, then the recoil valve switch can be turned on first, and then the recoil pump switch can be switched back on with air flow, until the air bubble closes and the recoil pump switch is turned off. Draw the valve switch, and then open the suction pump switch to continue suction filtration. After washing, close all valve switches and pump switches).
7. After all the six acids in the cooking tube are suctioned, the valve outlet on the rubber bottle of the distilled water bottle is inserted into the corresponding decontamination tube, and the valve on the rubber tube is opened. Distilled water is added to the decontamination tube and the tube is drained after 200 ml. , Continuously 2-3 times, until the test solution is used to test the filtered liquid to be neutral, then turn off the distilled water switch. The six boiling tubes were filtered with distilled water and tested with a dipstick.
8, then the alkaline rubber valve on the valve outlet into the appropriate cooking tube, open the valve on the rubber tube, respectively, in the decontamination tube to add 200ml of micro-boiling alkaline solution, close the valve, add in each decoction tube After the 2 drops of basal leaven, the alkali digestion, suction filtration and distilled water washing are performed according to the operation of the second half of Article 5 and the operation of 6th and 7th.
9, after the completion of the above work, use a straw to add 25ml or so of 95% B fermented yeast to the mouth of the decoction tube, soak it for 10 minutes after soaking.
10. Carefully pull up the lifting handle. At this time, the suction cup and the bowl will fall together (sometimes the cooking tube and the bowl will suck), and then the hand presses the suction filter base, so that the bowl and the suction base are loosened and lifted up. take out. (Alternatively, when the tube is decocted and sucked, turn the decontamination tube in a direction opposite to that of the bowl, so that the bowl is removed and removed).
11. Finally, remove the sealing ring from the crucible (see the structure diagram). Move the crucible into the incubator and dry it at 130°C±2°C for 2 hours. After removing it, cool it to room temperature in a desiccator and obtain the weight. M1. The weighed crucible was then placed in a high-temperature furnace at 500±25° C. for 1 hour, taken out and put in a desiccator and cooled to room temperature to obtain m2. The measurement result was calculated as follows:
M1― m2
Crude fiber%=DDDDDDDDDDDDD%m
Where: m1 ----- 130 °C after drying and sample residue weight;
M2-----500 °C after ignition and sample residue weight;
m------ sample (not degreased) quality.
12. After removing the switch, turn off the power

Polishing Grinding Test Equipment,Grinding Polishing Machine,Automatic Polish Grinding Test Equipment Manufacturers and Suppliers in China. Polishing Grinding Test Equipment,Grinding Polishing Machine,Automatic Polish Grinding Test Equipment,Metallographic Grinding Polishing MachineChina Polishing Grinding Test Equipment,Grinding Polishing Machine manufacturers, welcome Automatic Polish Grinding Test Equipment,Metallographic Grinding Polishing Machine purchasers from worldwide to visit our site.